• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Thu, 25.04.24

Search results


October 2017
Ron Sela MD, Mark Gellerman MD, Shaul Atar MD, and Eli Kalfon MD
June 2012
P. Codner, R. Nevzorov, J. Kusniec, M. Haim, R. Zabarski and B. Strasberg

Background: Defibrillation threshold (DFT) testing at the time of implantable cardioverter defibrillator (ICD) insertion is performed routinely. Recently this practice is being reconsidered due to doubts about its ability to improve ICD efficacy and evidence that survival may not be affected by the test.

Objectives: To compare the outcome of ICD recipients who underwent DFT testing and those in whom no testing was performed.

Methods: A total of 213 eligible patients were implanted with an ICD between 2004 and 2009. DFT testing was performed in 80. We compared total mortality, appropriate and inappropriate ICD shocks, and anti-tachycardia pacing (ATP) events between DFT and non-DFT patients during a follow-up of 2 years.

Results: On comparing the DFT and non-DFT groups, we found a 2 year mortality rate of 7.5% versus 8.3%, respectively (P = 0.8). Furthermore, 20.7% of patients in the DFT group and 12.4% in the non-DFT group had at least one episode of ICD shock (P = 0.15). With regard to ICD treatment (ICD shocks or ATP events), 57.7% in the DFT group and 64.2% in the non-DFT group received appropriate treatments (P = 0.78).

Conclusions: No significant differences in the incidence of 2 year mortality or percentage of ICD treatment emerged between the DFT and non-DFT groups.
 

Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
© All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel