A Novel Heart Rate Variability Algorithm for the Detection of Myocardial Ischemia: Pilot Data from a Prospective Clinical Trial

Click on the icon on the upper right hand side for the article written by Dan Oieru MD, Nir Shlomo, Israel Moalem, Eli Rozen MD, Alexey Naimushin MD, Robert Klempfner MD, Ilan Goldenberg MD and Ronen Goldkorn MD
IMAJ 2015: 17: March: 161-165
Abstract

Abstract



Background: Heart rate variability (HRV) analysis has been shown to be a predictor of sudden cardiac death and all-cause mortality in patients with cardiac disease.



Objectives: To examine whether newer HRV analysis algorithms, as used by the HeartTrends device, are superior to exercise stress testing (EST) for the detection of myocardial ischemia in patients without known coronary artery disease (CAD).



Methods: We present pilot data of the first 100 subjects enrolled in a clinical trial designed to evaluate the yield of short-term (1 hour) HRV testing for the detection of myocardial ischemia. The study population comprised subjects without known CAD referred to a tertiary medical center for EST with single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI). All patients underwent a 1 hour electrocardiographic acquisition for HRV analysis with a HeartTrends device prior to EST with MPI. Sensitivity, specificity, and positive and negative predictive values (PPV and NPV, respectively) were calculated for EST and HRV analysis, using MPI as the gold standard for the non-invasive detection of myocardial ischemia.



Results: In this cohort 15% had a pathologic MPI result. HRV analysis showed superior sensitivity (85%), PPV (50%) and NPV (97%) as compared to standard EST (53%, 42%, 90%, respectively), while the specificity of the two tests was similar (86% and 85%, respectively). The close agreement between HRV and MPI was even more pronounced among patients > 65 years of age.



Conclusions: Our pilot data suggest that the diagnostic yield of the novel HeartTrends HRV algorithm is superior to conventional EST for the non-invasive detection of myocardial ischemia.


Download PDF Print
Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
© All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel